Biomechanical comparison of different interspinous process devices in the treatment of lumbar spinal stenosis: a finite element analysis

文献详情

资源类型:
WOS体系:
Pubmed体系:

收录情况: ◇ SCIE

单位: [1]Capital Med Univ, Beijing Friendship Hosp, Dept Orthopaed, Beijing 101100, Peoples R China [2]Chengde Med Univ, Dept Spine Surg, Affiliated Hosp, Chengde 067000, Hebei, Peoples R China [3]Chengde Med Univ, Dept Joint Surg, Affiliated Hosp, Chengde 067000, Hebei, Peoples R China
出处:
ISSN:

关键词: Lumbar spinal stenosis Finite element analysis Biomechanics Interspinous process devices BacFuse Coflex X-Stop

摘要:
Background Lumbar spinal stenosis (LSS) is a common disease among elderly individuals, and surgery is an effective treatment. The development of minimally invasive surgical techniques, such as the lumbar interspinous process device (IPD), has provided patients with more surgical options. Objective To investigate the biomechanical properties of different IPDs, including BacFuse, X-Stop and Coflex, in the treatment of LSS. Methods Based on the computed tomography images of a patient with LSS, four finite element (FE) models of L3-S5 were created in this study. The FE models included a surgical model of the intact lumbar spine and surgical models of the lumbar IPDs BacFuse, X-Stop, and Coflex. After validating the models, they were simulated for four physiological motions: flexion, extension, lateral bending and axial rotation, and range of motion (ROM). Stress distribution of discs and facet joints in each segment, stress distribution of the spinous process in the operated section, and stress distribution of the internal fixation were compared and analysed. Results Compared to the model of the intact lumbar spine, the other three models showed a decrease in ROM and disc and facet joint stresses in the surgical segment during movement and an increase in ROM and disc and facet joint stresses in the adjacent segments. These effects were greater for the proximal adjacent segment with BacFuse and more pronounced for the distal adjacent segment with Coflex, while X-Stop had the greatest stress effect on the spinous process in the surgical segment. Conclusion BacFuse, Coflex and X-Stop could all be implemented to effectively reduce extension and disc and facet joint stresses, but they also increase the ROM and disc and facet joint stresses in adjacent segments, which may cause degeneration.

语种:
WOS:
PubmedID:
中科院(CAS)分区:
出版当年[2021]版:
大类 | 3 区 医学
小类 | 4 区 骨科 4 区 风湿病学
最新[2025]版:
大类 | 3 区 医学
小类 | 3 区 骨科 3 区 风湿病学
JCR分区:
出版当年[2020]版:
Q3 ORTHOPEDICS Q4 RHEUMATOLOGY
最新[2023]版:
Q2 ORTHOPEDICS Q3 RHEUMATOLOGY

影响因子: 最新[2023版] 最新五年平均[2021-2025] 出版当年[2020版] 出版当年五年平均[2016-2020] 出版前一年[2019版] 出版后一年[2021版]

第一作者:
第一作者单位: [1]Capital Med Univ, Beijing Friendship Hosp, Dept Orthopaed, Beijing 101100, Peoples R China [2]Chengde Med Univ, Dept Spine Surg, Affiliated Hosp, Chengde 067000, Hebei, Peoples R China
通讯作者:
推荐引用方式(GB/T 7714):
APA:
MLA:

资源点击量:1320 今日访问量:0 总访问量:816 更新日期:2025-04-01 建议使用谷歌、火狐浏览器 常见问题

版权所有:重庆聚合科技有限公司 渝ICP备12007440号-3 地址:重庆市两江新区泰山大道西段8号坤恩国际商务中心16层(401121)