AIM: To compare the differences and agreement of ocular biometric parameters in highly myopic eyes obtained by optical biometric measurement instruments, the OA-2000 and IOLMaster 500. METHODS: Totally, 90 patients (90 eyes) were included. They were divided into high myopia group and control group. Ocular parameters, including axial length (AL), mean keratometry (Km), anterior chamber depth (ACD), and white to white (WTW), were obtained from the OA-2000 and IOLMaster 500. RESULTS: For the control group, we applied Bland-Altman graphs to assess the 95% limits of agreement (LoA) for most parameters including AL, ACD, Km, and WTW (-0.24 to 0.29 mm, -0.22 to 0.45 mm, -0.39 to 0.31 D, and -0.90 to 0.86 mm, respectively). In high myopia patients, AL, ACD, Km values had wider 95% LoA (-0.34 to 0.32 mm, -0.36 to 0.34 mm, -0.57 to 0.47 D, respectively), except WTW (-0.80 to 0.68 mm). Differences were not statistically significant between these two instruments (P>0.05). CONCLUSION: Most parameters obtained by the OA-2000 and IOLMaster 500 are comparable, including the AL, ACD, and K values. Among them, the agreement of the high myopia patients is poor compared to the patients without high myopia.
基金:
National Natural Science Foundation of ChinaNational Natural Science Foundation of China (NSFC) [81870686]; Beijing Municipal Natural Science FoundationBeijing Natural Science Foundation [7184201]; Capital's Funds for Health Improvement and Research
第一作者单位:[1]Department of Ophthalmology, Beijing Friendship Hospital, Capital Medical University, Beijing 100050, China
通讯作者:
通讯机构:[1]Department of Ophthalmology, Beijing Friendship Hospital, Capital Medical University, Beijing 100050, China[*1]Department of Ophthalmology, Beijing Friendship Hospital, Capital Medical University, 95 Yong-an Rd, Beijing 100050, China
推荐引用方式(GB/T 7714):
Guo Xiao-Xiao,You Ran,Li Shan-Shan,et al.Comparison of ocular parameters of two biometric measurement devices in highly myopic eyes[J].INTERNATIONAL JOURNAL of OPHTHALMOLOGY.2019,12(10):1548-1554.doi:10.18240/ijo.2019.10.05.
APA:
Guo, Xiao-Xiao,You, Ran,Li, Shan-Shan,Yang, Xiu-Fen,Zhao, Lu...&Chen, Xi.(2019).Comparison of ocular parameters of two biometric measurement devices in highly myopic eyes.INTERNATIONAL JOURNAL of OPHTHALMOLOGY,12,(10)
MLA:
Guo, Xiao-Xiao,et al."Comparison of ocular parameters of two biometric measurement devices in highly myopic eyes".INTERNATIONAL JOURNAL of OPHTHALMOLOGY 12..10(2019):1548-1554